Specification:
Model |
ACD-8BL |
Description |
Small Autoclave |
Class |
B |
Standard |
BS EN 13060:2004 + A2:2010 |
Frequency |
50Hz |
Plug |
3 hole plug |
Wattage |
2.4 KW |
Sterilization temperature |
105 - 134℃ |
Max working pressure |
0.21-0.23MPA |
Diameter of chamber |
17 CM |
Depth of chamber |
32 CM |
Volume of chamber |
8 L |
Material of chamber |
Stainless steel 304 |
Colors available |
White |
Screen type |
LCD |
Screen size |
6 X 4.25 cm |
Language available |
English, Italian, German |
Lock system of door |
Mechanic + electrical lockers |
User password |
Yes |
Output |
Thermo printer & USB |
Net weight |
35 kgs |
Gross weight / carton |
42 kgs |
Product size |
56 X 44 X 36cm |
Package size |
65 X 45 X 42cm |
Min size of table for installation |
54 x 41.5 cm |
Funication:
Fast Sterilization
Fully Automatic
- 8 Preset sterilization programs
- Ster. Temp.: 105-134℃
- Ster. Time: 4-60 min
- Dry time: 1-25min
- Vacuum Times: 1-10 times
- B&D / Helix / Vacuum testing program
- Cleaning Program
- Printer & USB Output
- Big Stainless Steel Water Tanks
- High quality bacteria filter
User Password
Program |
121℃ Solid |
121℃ Hollow |
121℃ Porous |
121℃ Solid |
121℃ Hollow |
121℃ Porous |
Vacuum Times |
1 |
3 |
3 |
1 |
3 |
3 |
Temp (℃) |
121 |
121 |
121 |
134 |
134 |
134 |
Pressure(KPA) |
110 |
110 |
110 |
210 |
210 |
210 |
Time(Min.) |
20 |
20 |
20 |
4 |
4 |
4 |
Program |
Cotton |
Prion |
User defined |
BD/Helix Test |
Vacuum Test |
Clean |
Vacuum Times |
3 |
3 |
User defined |
3 |
1 |
3 |
Temp (℃) |
134 |
134 |
User defined |
134 |
/ |
105 |
Pressure(KPA) |
210 |
210 |
User defined |
210 |
-80 |
20 |
Time(Min.) |
4 |
19 |
User defined |
3.5 |
15 |
5 |