Model :MS500
Features:
The MS500, a new type full - spectrum ( 200~ 850nm) spectrophotometer, can measure nucleic acids, proteins, cell cultures, microarray samples, and UV-VIS full spectrum absorbance scan with high accuracy and reproducibility.
Specification:
sample size |
0.3~2µL |
Rapid measure |
speed |
Display the concentration directly |
|
Performance index: |
|
Path length |
1mm, 0.2mm, (0.1mm, 0.05mm optional) |
Sample size |
0.3~2µL |
Light source |
Xenon flash lamp |
Detector |
2048-element linear silicon CCD |
Wavelength range |
200~850nm |
Wavelength accuracy |
±1nm |
Wavelength resolution |
2nm (FWHM at Hg 546nm) |
Absorbance precision |
0.002 Abs |
Absorbance accuracy |
1% (0.76 absorbance at 350nm) |
Absorbance range |
0.02~75(150, 300 optional, 10mm equivalent absorbance) |
Detects Nucleic acid up to |
2~3750ng/μL(7500,15000 optional, dsDNA) |
Detects Protein up to |
0.1mg/ml~100mg/ml(200,400 optional, BSA) |
Measurement time |
<5s |
Outline size |
21cm×17cm×11cm |
Net weight |
1.35kg |
No cells, no cuvettes ,No warm-up |
|
UV-VIS full spectrum absorbance analysis (200-850nm) |
Model ;MS600
MS600 is a new type full – spectrum ( 200~ 850nm) spectrophotometer, can measure nucleic acids, proteins, cell cultures, microarray samples, and UVVIS full spectrum absorbance scan with high accuracy and reproducibility. Adding dark room for cuvette, you can choose cuvette or test platform
Specification :
sample size |
0.3~2µL |
Cuvette or test platform |
|
No warm-up |
|
Display the concentration directly |
|
Performance index |
|
Path length |
1mm, 0.2mm, (0.1mm, 0.05mm optional) |
Sample size |
0.32µL |
Light source |
Xenon flash lamp |
Detector |
2048-element linear silicon CCD |
Wavelength range |
200850nm |
Cuvette size(path length) |
1mm, 2mm, 5mm, 10mm |
Wavelength accuracy |
±1nm |
Wavelength resolution |
2nm (FWHM at Hg 546nm) |
Absorbance precision |
0.002 Abs |
Absorbance accuracy |
1% (0.76 absorbance at 350nm) |
Absorbance range |
0.002~75(150, 300 optional, 10mm equivalent absorbance) |
Detects Nucleic acid up to |
0.4~3750ng/μL(7500,15000 optional, dsDNA) |
Detects Protein up to |
0.01mg/ml~100mg/ml(200,400 optional, BSA) |
Measurement time |
<5s |
Outline size |
24cm×21cm×11cm |
Net weight |
1.92kg |
UV-VIS full spectrum absorbance analysis (200-850nm) No dilutions for samples, can measure samples with high concentration!Easy to use |
General operation
1 Drop directly 0.32µL sample onto the test platform with a pipette.
2 Close the sampling arm.
3 Or open the dark room and put the cuvette into it.
4 Click the “Measure” button, and the result(values and graph