Availability: In Stock
Product Code: 955
Brand: Gas chromatography

GC12A

$0

                    Characteristics:

  • Brand-new designed microcomputer temperature control system of world advanced control technology. High accuracy of temperature control (optimum: ± 0.05OC), high reliability and anti-interference.
  • Control and operating parameters (including for detectors) set up by the keyboard. Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting. Accurate display of setting and actual value of temperature control of each path, bridge current of TCD, sensitivity of FID, retention and analysis time.
  •  Five paths of independent control sections: column oven, sample injector, detector and thermal conductive pond.
  • 5-ramp programmed temperature rising control. Auto-drop of temperature of column oven (auto-opening the door behind the oven) to a fast cooling.
  •  transformation oven, absorber and pyrolysis absorber available for selection. Two detectors can be installed at the same time with high sensitivity. Short time to be stabilized and easy to clean and install nozzle.
  •  Packed column and capillary column (0.53 large diameter or small diameter) can be installed easily including a complete set of split sampling connector and tailing connector to be installed free of charge. Optional 0.53mm large diameter capillary direct sampler is available for choice.
  •  Analysis can be conducted in packed or capillary column with multiple sampling methods.
  • Packed column: on-column sampling, instantaneous vaporized sampling and gas sampling.
  • Capillary column: split sampling and 0.53mm large diameter capillary direct sampling.
  • Construction of dual gas path, dual sample injector and dual packed column system can effectively avoid baseline undulation and drift in real time and facilitate programmed temperature rising analysis. Accurate and scaled type gas path control valves will provide a high repeatability and stability.
  • Advanced functions: auto ON/OFF temperature control system, permanent data storage, file storage and call and over-temperature protection.
  • Open-end computer system can apply RS-232 interface and chromatogram workstation to achieve dual path communication control and data processing.
  • Large volume column (300x280x270) can hold capillary column and dual packed column. Built-in heating filament structure is in favor of quartz capillary analysis.
  • A complete set of accessories including purifier, pressure-reducing valve, gas path tube, wrench and tools, injection needle and connectors of all kinds are provided free of charge and are ready to connect with supply-air system prepared by users.
  • CAD design, mechanical parts process based on the imported HITACHI CNC processing center and AMADA CNC punch press and other advanced equipment, It has high precision, structural reliability, good looks and hommization

                 Specification:

1/sample volume range

gas 0.01ml- 2ml; liquid 0.01μl - 10μl

2/ Temperature control range of columns box,injectoranddetector

room temperature plus 7 °C - 400 °C

3/Optimum testing result of flame ionization detector (FID)

≤ 1 × 10-11g / s (sample: n-hexadecane)

4/Sensitivity of Thermal conductivity detector (TCD)

 ≥ 2500mV • ml / mg (sample: hexadecane, carrier gas: H2)

5/dimensions l × b × h

640 mm × 485 mm × 465 mm

6/mass (weight)

70kg

Related Products